Noise Measurement  [Home] [Archive] [Noise]
Shielding Sound Level (dBA) Test Track Spectrogram Typical Power Spectrum
Room Background ep2d_diff, in bore ep2d_diff, off bore
No shielding 63 109 109 -35dB from original sound level Click to View Click to View
Peak frequency about 640Hz
AVOTEC Full Protection Ear Set Closed chamber 39 80 80 -10dB from original sound level Click to View Click to View
Peak frequency about 620Hz
Over ear 96 -18dB from original sound level Click to View Click to View
Peak frequency about 625Hz
3M1440 Ear Muff Converted Ear Set Closed chamber 38 76 74.5 -3dB from original sound level Click to View Click to View
Peak frequency about 620Hz
Over ear 94.2 -15dB from original sound level Click to View Click to View
Peak frequency about 625Hz
Pressed Foam Closed chamber 58 107.6 107
Over ear 107.8
fmri
No shielding 105 -40dB from original sound level Click to View Click to View
Peak frequency about 640Hz
3M1440 Ear Muff Converted Ear Set Closed chamber 73.6 Original sound level Click to View Click to View
Peak frequency about 745Hz
Over ear 90.8 -3dB from original sound level Click to View Click to View
Peak frequency about 620Hz