Auto-Calibrating the Z-shim for Slice Specific
Recovery of Susceptibility Losses
Keith Heberlein & Xiaoping Hu
ISMRM-2005 #2271
|
|
|
|
Background
Z-shim is a method for recovering susceptibility related signal
loss in T2*-weighted images.
Several pulse sequences incorporate the Z-shim for fMRI studies
investigating problematic areas of the brain such as amygdala and
orbital frontal cortex humans. [1-5]
Essentially, these brain regions are affected by local field
present across the slice.
This gradient field causes spin dephasing resulting in signal
loss.
The dephasing can be refocused using a small compensation gradient
called the Z-shim.
Calibrating the Z-shim manually for each slice is time consuming
and operator dependent.
Using a field map and a linear approximation for the field gradient
across the slice, the Z-shim can be calibrated automatically for each
slice.
Methods


Fig.1 Field Map to Gradient Map
Field maps [Fig. 1] are obtained using spin echo EPI on a 3T Siemens Trio MR
Scanner using two acquisitions, one with a 2 ms echo offset.
The field map is masked and median filtered to remove spurious
pixel values.
At each pixel the distortion gradient is approximated by a linear
fitting across three adjacent slices.
The z-shim is estimated by the mean distortion gradient in the
strongest region of susceptibility artifact and the requested echo
time.
The method is demonstrated in vivo using the Z-SAGA
[1] pulse sequence, but is generally applicable to most z-shim
methods.
|
The field map is masked and median filtered to remove spurious
pixel values.
At each pixel the distortion gradient is approximated by a
linear fitting across three adjacent slices.
The z-shim is estimated by the mean distortion gradient in the
strongest region of susceptibility artifact and the requested
echo time [fig. 2].

Fig. 2 In Vivo Image Profile of Field Gradient Distortion
The method is demonstrated in vivo using the Z-SAGA
[1] pulse sequence, but is generally applicable to most
z-shim methods.
Results
Fig. 3 Z-shim result
Discussion
The auto-calibration procedure is effective in determining slice
specific z-shim values.
Extending the method to include more than a single z-shim is
possible as in Ref. 5, but at the expense
of temporal resolution and/or slice coverage.
Acknowledgment
This work was supported in part by the National Institutes of
Health, Georgia Research Alliance and the Whitaker Foundation.
|
References
- Heberlein KA, Hu X. Simultaneous acquisition of gradient-echo and asymmetric spin-echo for single-shot z-shim: Z-SAGA. Magn Reson Med 2004;51(1):212-216
- Guo H, Song AW. Single-shot spiral image acquisition with embedded z-shimming for susceptibility signal recovery. J Magn Reson Imaging 2003;18(3):389-395
- Li Z, Wu G, Zhao X, Luo F, Li SJ. Multiecho segmented EPI with z-shimmed background gradient compensation (MESBAC) pulse sequence for fMRI. Magn Reson Med 2002;48(2):312-321
- Gu H, Feng H, Zhan W, Xu S, Silbersweig DA, Stern E, Yang Y. Single-shot interleaved z-shim EPI with optimized compensation for signal losses due to susceptibility-induced field inhomogeneity at 3 T. Neuroimage 2002;17(3):135-136
- Deichmann R, Josephs O, Hutton C, Corfield DR, Turner R. Compensation of susceptibility-induced BOLD sensitivity losses in echo-planar fMRI imaging. Neuroimage 2002;15(1):120-135
|
|